Previous Next
Im Interested in this Product. Submit Inquiry

EXCELSIOR

Last Updated on : 10-1-2017

EXCELSIOR

VAPORIZER MANIFOLD

Provision for mounting 2 temperature and flow compensated vaporizers.

FLOWMETER MANIFOLD

  • 5 Tube flowmeter set
  •  Each high & Low range tube set for O2 & N2O 
  • 1 High range tube for AIR
  • Each gas control value fitted with colour and shape control knows.
  • Fitted with oxygen ratio controller which ensure minimum 25% O2 concentration. If oxygen delivery is altered. ORC automatically readjusts proportionate delivery of N2O, thereby maintaining pre-set oxygen concentration.
 Gas Flowmeter  Range 
 O2 Double 

0 to 1 Litre/min

1 to 12 Litre/min   

 N2O Double 

0 to 1 Litre/min

1 to 12 Litre/min 

 AIR single 

100 to 15000 cc/min 

      GAS INLET

  • Cylinder inlet : 4 forget brass pin indexed yokes, 2 each for O2 & N2O with colour coded cylinder pressure gauge on the front at eye level.
  • Pipeline inlet : Non-interchangeable pipeline inlet for O2 N2O & AIR with separate pipeline pressure gauges on the font at eye level.
 Application Gas  Qty.  Pressure gauge range 
 cylinder O2  0-250 kg/cm2 
 cylinder N2O  0-250 kg/cm2 
 Pipe line O2   1 0-7 kg/cm2 
 Pipe line  N2O  0-7 kg/cm2 
 Pipe line  AIR  0-7 kg/cm2 
 Vacuum system VACUUM 

0-760 mm/Hg 

PRESSURE REGULATORS

4 first stage regulators, 2 each for O2 & N2O. 2 second stage regulators for maintaining systems pressure. 1 each for O2 & N2O   

 

Description