Previous
Next
Im Interested in this Product. Submit Inquiry
EXCELSIOR
Last Updated on : 10-1-2017
EXCELSIOR
VAPORIZER MANIFOLD
Provision for mounting 2 temperature and flow compensated vaporizers.
FLOWMETER MANIFOLD
- 5 Tube flowmeter set
- Each high & Low range tube set for O2 & N2O
- 1 High range tube for AIR
- Each gas control value fitted with colour and shape control knows.
- Fitted with oxygen ratio controller which ensure minimum 25% O2 concentration. If oxygen delivery is altered. ORC automatically readjusts proportionate delivery of N2O, thereby maintaining pre-set oxygen concentration.
Gas | Flowmeter | Range |
O2 | Double |
0 to 1 Litre/min 1 to 12 Litre/min |
N2O | Double |
0 to 1 Litre/min 1 to 12 Litre/min |
AIR | single |
100 to 15000 cc/min |
GAS INLET
- Cylinder inlet : 4 forget brass pin indexed yokes, 2 each for O2 & N2O with colour coded cylinder pressure gauge on the front at eye level.
- Pipeline inlet : Non-interchangeable pipeline inlet for O2 N2O & AIR with separate pipeline pressure gauges on the font at eye level.
Application | Gas | Qty. | Pressure gauge range |
cylinder | O2 | 2 | 0-250 kg/cm2 |
cylinder | N2O | 2 | 0-250 kg/cm2 |
Pipe line | O2 | 1 | 0-7 kg/cm2 |
Pipe line | N2O | 1 | 0-7 kg/cm2 |
Pipe line | AIR | 1 | 0-7 kg/cm2 |
Vacuum system | VACUUM | 1 |
0-760 mm/Hg |
PRESSURE REGULATORS
4 first stage regulators, 2 each for O2 & N2O. 2 second stage regulators for maintaining systems pressure. 1 each for O2 & N2O