Previous Next
Im Interested in this Product. Submit Inquiry

SELECTATEC FITTING WITH INTERLOCK

Last Updated on : 10-1-2017

SELECTATEC FITTING WITH INTERLOCK

 DIMENSION  Height Width
Depth
Weight (Approx)
Agent capacity retention by wicks 60 ml

SELECTATEC with interlock

 235mm 114mm
220mm
7.35Kg
225ml +- 25ml
 cagemount (iso 23mm tapers)  238mm 137mm
200mm
7.00 Kg

225ml +-25ml

 OUTPUT ACCURACY :

5 LPM O2 @ 21C +- 2C

Vaporizer 5% : +-25% of the delivered agent or +-15% of the dial setting (whichever is greater)

Vaporizer 8% : +-4 % of the delivered agent or +-15% of the dial setting (whichever is greater)

calibrated at sea level

PERFORMANCE RELIABILITY

    Performs with gas flow rates from 0.2 - 15 lpm

ENVIRONMENTAL

TEMPERATURE : 15C to 35C

HUMIDITY : 30% to 75% relative humidity

works consistently in altitudes up to 2440m asi

VAPORISER TYPE

  Halothane 5%, 6%, 8% filler type 
sevoflurane 8% filler type
 isoflurance 5%,6%,8% filler type
Enflurane 5% 7% filler type
   screw  keyed
 screw keyed quik
 screw  keye
 screw  keyed
selectatec compatible with interlock
 TRUE  TRUE
 TRUE TRUE
 cagemount (iso 23mm tapers)
 TRUE  TRUE  TRUE  TRUE
 Drager plug in
 TRUE  TRUE  TRUE  TRUE

 

 

Description