Specialist Of Sleep Diagnosis And Therapy
Last Updated on : 10-1-2017
Specialist of sleep diagnosis and therapy
- Technical Specifications
- Device size
- Dimension : 220*194*112mm 313*194*112mm (with inH2 heated humidifier)
- Weight : < 2.2kg <3 kg (with inH2 heated humidifier)
Environmental
operating | Transport and storage | |
Temperature | 5 to 30c | -20 to 55c |
Relative hurnicity | < 80% Non-condensing | < 93% Non-condensing |
Atmospheric Pressure | 860 to 1060 hpa | 500 to 1060 hpa |
STANDARDS COMPLIANCE
- IEC 60601-1 general requirements for safety of medical electrical equipment
- IEC 60601-1-2 electromagnetic compatibility
- ISO 17510-1 sleep apnoea breaking therapy
- ISO 8185 general requirements for humidification system
ELECTRICAL
AC power consumption : 100-240v AC, 50/60Hz, Max 95VA Type of protection against electric shock : class 2 equipment degree of protection against electric shock : Type BF applied part degree of protection against ingress of water : Device : Drip proof, IPX1 mode of operation : continuous
MODE COMPARISON
Product Model |
Pressure |
Function |
Auto Pressure |
Sleepsmart BPAP 25 |
4.25cm H2O |
S, CPAP |
No |
Sleepsmart BPAP 25A |
4.25cm H2O | S,CPAP |
available |
Sleepsmart BPAP 25 T |
4.25cm H2O | S,CPAP, T,S/T |
No |
Sleepsmart BPAP 30 T |
4.30cm H2O | S,CPAP,T,S/T |
No |
- IPAP : 4 to 30 cmH2O +-1 cmH2O**
- EPAP : 4 to 25 cmH2o +-1 cmH2O**
- CPAP : 4 to 20 cmH2o +-1 cmH2O**
- Respiratiory Rate : 3 to 30 BPM, greater of +-1 BPM or +-10% of the setting (when measured over a 4 minutes period)
- E/I Rate : 1-4
- Max. inspiration time : 3.0 seconds
- Ramp. duration : 0 to 45 minutes +-10% of the setting
- Rise time : 1 to 3*** +- 25%*****
- Limited to 20 cmH2o when using auto feature in s mode
- Dynamic pressure accuracy is +-1 cmH2O measured at the patient end of the circuit with a vio nasal mask and varying flow conditions.
- Static pressure accuracy is +-0.5 cmH2O measured at the patient end of the circuit with a vio nasal mask and no patient flow.
- The range of values corresponds to tenths of seconds (e.g.., a setting of 4 indicates a rise time of 0.4 seconds)
- Measured at the patient end of circuit with a vio nasal mask exhalation device and no patient flow.
DISPOSAL
Dispose of the device in accordance with local regulations.
NOTE :
Specifications subject to change without prior notice